Serviço
de Assessoria Científica - 0800-315454
|
PROGRAMAÇÃO
PARA SELECTRA II e E, XL e FLEXOR ®
| ||||
|
TRIGLICÉRIDES
LÍQUIDO ESTÁVEL - CRYSTAL
| ||||
|
Catálogo:
|
K.055-1
/ K.55-2
|
Número
de Testes:
|
333 / 666
| |
Preparação
e estabilidade do reagente de trabalho: utilizar as instruções de uso.
TEST PARAMETER MONO/DUAL MODE – 37 °C
| Name |
TRIGLICÉRIDES
|
Prozone check |
No
|
| Abbr.Name |
TRIG
|
Ref. Male low |
*
|
| Mode |
Endpoint
|
Ref. Male high |
*
|
| Wavelength |
505
nm
|
Ref. Female high |
*
|
| Units |
mg/dL
|
Ref. Ped low |
*
|
| Decimals |
0
|
Ref. Ped high |
*
|
| Low Conc. |
0
mg/dL
|
Control 1 |
BIOCONTROL
|
| High Conc. |
900
mg/dL
|
- |
-
|
| Calibrator Name |
BIOCAL
|
Control 2 |
*
|
| Repeat |
2
|
-- |
-
|
| Number |
1
|
Control 3 |
*
|
| Concentration |
*
|
- |
-
|
| Interval |
0
|
Correlat factor |
1.000
|
| Cut off |
No
|
Correlat offset |
0.000
mg/dL
|
| DUAL MODE |
.
|
. |
.
|
| Name |
TRIGLICÉRIDES
|
-- |
-
|
| Sample blank |
No
|
Low Absorbance |
-0.100
Abs
|
| R1 bottle |
25 mL
|
High Absorbance |
3.000
Abs
|
| Normal volume |
300
µL
|
R.Abs.L.Limit |
-0.100
Abs
|
| Rerun volume |
301 µL
|
R.Abs.H.Limit |
3.000
Abs
|
| Sample |
-
|
- |
-
|
| Normal volume |
3
µL
|
Reagent Blank |
Yes
(#)
|
| Rerun volume |
2
µL
|
Cal Low Limit |
-0.100
|
| R2 bottle |
5
mL
|
Cal High Limit |
3.000
|
| Normal volume |
0
µL
|
- |
-
|
| Rerun volume |
0
µL
|
Factor |
#
|
| Predilution |
No
|
- |
-
|
| Incubation time |
11.5
min
|
- |
-
|
| MONO MODE |
.
|
. |
.
|
| Name |
TRIGLICÉRIDES
|
- |
-
|
| Sample blank |
No
|
Low Absorbance |
-0.100 Abs
|
| R1 bottle |
25
mL
|
High Absorbance |
3.000
Abs
|
| Normal volume |
300
µL
|
R.Abs.L.Limit |
-0.100 Abs
|
| Rerun volume |
301 µL
|
R.Abs.H.Limit |
0.300 Abs
|
| Sample |
-
|
- |
-
|
| Normal volume |
3
µL
|
Reagent Blank |
Yes
(#)
|
| Rerun volume |
2
µL
|
Cal Low Limit |
*
|
| - |
-
|
Cal High Limit |
*
|
| Incubation Time |
11.5
min
|
Factor |
#
|
*Selecionado
pelo usuário. .....#Calculado pelo analisador
Ver
Instruções de Uso
Rev. Mar/2010